Double lumen tube
Adv:
- quicker to insert
- less prone to displacement
- can accommodate the deflation & suctioning of secretions from either lung
Features & design
Modern DLTs are single use, made of polyvinylchloride and are based on the Robertshaw design
DLT sizes are sized using the French (Fr) gauge system, where 1 Fr=1/3 mm of the outer diameter. Thus, a 37 Fr DLT has an outer diameter of 1/3×37=12.3 mm.
DLT vs BB
Sequence of clinical check

- Correct bronchus
- Clamp tracheal lumen
- ventilate bronchial lumen
- → should ventilate left lung
- if right lung ventilates → DLT in RMB
- Correct depth
- Clamp bronchial lumen
- ventilate tracheal lumen
- → right lung easily ventilated
- if ↑Paw: tracheal lumen obstructed by wall of LMB
- → need shift out
- if both lungs ventilate → too far out
- → need shift in
- if ↑Paw: tracheal lumen obstructed by wall of LMB
- Bronchoscopy